5.2.3. Setup
5.2.4. Test Procedure
The test procedure was in accordance with EN 61000-4-3
a) The testing was performed in a fully anechoic chamber. The transmit antenna was located at a
distance of 3 meters from the EUT.
b) The frequency range is swept from 80 MHz to 1000 MHz, with the signal 80% amplitude modulated
with a 1kHz sine-wave. The rate of sweep did not exceed 1.5 x 10 -3 decade/s, where the frequency
range is swept incrementally, the step size was 1% of preceding frequency value.
c) The dwell time at each frequency shall be not less than the time necessary for the EUT to be able to
respond.
d) The test was performed with the EUT exposed to both vertically and horizontally polarized fields on
each of the four sides.