3.4. Test Site Environment
Items Test Item Required (IEC 60068-1) Actual
Temperature (°C) 15-35 26.0
Humidity (%RH) 25-75 60.0
Barometric pressure (mbar)
EN 55022 CE
860-1060 950
Temperature (°C) 15-35 26.0
Humidity (%RH) 25-75 60.0
Barometric pressure (mbar)
EN 55022 RE
860-1060 950
Temperature (°C) -- 26.0
Humidity (%RH) -- 60.0
Barometric pressure (mbar)
EN 61000-3-2
-- 950
Temperature (°C) -- 26.0
Humidity (%RH) -- 60.0
Barometric pressure (mbar)
EN 61000-3-3
-- 950
Temperature (°C) 15-35 24.0
Humidity (%RH) 30-60 43.6
Barometric pressure (mbar)
EN 61000-4-2
860-1060 950
Temperature (°C) -- 20.1
Humidity (%RH) -- 54.2
Barometric pressure (mbar)
EN 61000-4-3
-- 950
Temperature (°C) 15-35 22.3
Humidity (%RH) 30-60 54.8
Barometric pressure (mbar)
EN 61000-4-4
860-1060 950
Temperature (°C) 15-35 24.0
Humidity (%RH) 10-75 52.1
Barometric pressure (mbar)
EN 61000-4-5
860-1060 950
Temperature (°C) -- 24.0
Humidity (%RH) -- 52.1
Barometric pressure (mbar)
EN 61000-4-6
-- 950
Temperature (°C) 15-35 22.6
Humidity (%RH) 25-75 51.6
Barometric pressure (mbar)
EN 61000-4-8
860-1060 950
Temperature (°C) 15-35 19.1
Humidity (%RH) 25-75 54.6
Barometric pressure (mbar)
EN 61000-4-11
860-1060 950
A Test Lab Techno Corp.
Report Number: 1402CE13-01
16 of 153